Graph explanation
1) Graph 1 is to compensate for pressure loss in piping between the ejector and the pressure

gauge, and find Po (pressure setting).

2) Graph 2 shows correlation between P (drive pressure) and Qa (air consumption).
3) Graph 3 shows correlation between P (drive pressure) and Pg (suction pressure; when the

sample gas inlet of the ejector is closed).

4) Graph 4 shows correlation between P (drive pressure) and Qg (suction flow) for each gas

pressure.

Qa

L

200

100

0

5

10

15

P= 0.5

0

40

30

40

60

80

Po (kPa)

Qa (l/min)

0

-1.0

-0.5

40

60

80

Pg (kPa)

0

8

4

40

60

80

Qg (l/min)

L (m)

P (kPa)

P (kPa)

P (kPa)

Po (kPa) : Pressure setting

P (kPa) : Drive pressure (at the ejector entrance)

Pg (kPa) : Suction pressure

Qa (l/min) : Air consumption

Qg (l/min) : Suction flow

L (m) : Distance between the ejector and the pressure

gauge

Needle

valve

Pressure gauge

Ejector

Po

Air

source

Gas

Pg

Qg

Gas pressure : -15 Pa

Air consumption characteristics

Pressure setting characteristics

Suction pressure characteristics

Suction flow characteristics

F2-9E.ai

Gas Pressure:

0 kPa

1)

2)

3)

4)

Needle valve

<1>

<2>

Pressure gauge

Pressure gauge assembly

Nozzle (Note1)

R1/2

Ejector

Detector

<1> Rc1/4 or 1/4 FNPT

<2> Ш6/Ш4 or 1/4 inch copper tube (stainless)

with ejector to connect

<3> R1/4 or 1/4 NPT

40

20

Tee

Blow Rc1/4

Ø43

Approx. 67

Approx. 88

39

Approx. 70 38

Full open height

(Note1) The connecter of ejector assembly is a

dedicated connecter with nozzle function.

Instrument air inlet