6 confirmation of detector type setting, 7 selection of sample gas, 8 output range setting – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 96: Confirmation of detector type setting -8, Selection of sample gas -8, Output range setting -8 7.8.1, Minimum current (4 ma) and maximum current (20 ma), Settings -8, Caution, Ma-output1 range

Advertising
background image

<7. Startup>

7-8

IM 11M12A01-02E

7.6

Confirmation of Detector Type Setting

Check that the detector in Figure 7.7 is the one for this equipment.

CAUTION

• If this converter is to be used in conjunction with the ZO21D, the power requirements are

limited to 125 V AC or less, 50 Hz or 60 Hz (it cannot be used with a 125 V or greater, or in

the EEC).

• If detector settings are to be changed, first disconnect the wiring connections between the

detector and the converter. Then change detector settings appropriately.

7.7

Selection of Sample Gas

Combustion gases contain moisture created by burning hydrogen in the fuel. If this moisture is

removed, the oxygen concentration might be higher than before. You can select whether the

oxygen concentration in a wet gas is to be measured directly, or compensated for its dry-gas

value before use. Select the “Select measure gas: Wet” in Figure 7.8 to select either wet or dry

gas.

7.8

Output Range Setting

This section sets forth analog output range settings. For details, consult Section 8.1, Current

Output Setting, later in this manual.

7.8.1

Minimum Current (4 mA) and Maximum Current (20

mA) Settings

To set the minimum and maximum current settings, follow these steps:
(1) Select the “Commissioning” from the Execution/Setup display.
(2) From the Commissioning (Setup) display, select “mA-output setup”; the display shown in

Figure 7.9 then appears.

(3) Select “mA-output1” from the mA-outputs display. The “mA-output1 range” display shown in

Figure 7.10 then appears.

(4) In the display shown in Figure 7.10, select “Min. oxygen conc” and press the [Enter] key to

display the numeric-data entry display. Enter the oxygen concentration at a 4-mA output;

enter [010] for a ten-percent oxygen concentration measurement.

(5) Also in Figure 7.10, select “Max. oxygen conc” at a 20-mA output. Enter the appropriate

maximum oxygen concentration (at the 20-mA output) in the same manner as in step 4

above.

(6) Set “mA-output2” in the same manner as in the appropriate steps above.

Both outputs are 4-20mA
mA-output1

mA-output2

Set presets

Enter

mA-outputs

F7.9E.ai

Parameter: Oxygen

Set range

Min. oxygen conc:

0 %O

2

Max. oxygen conc:

25 %O

2

Output damping: 0 s

Output mode: Linear

Enter

mA-output1 range

F7.10E.ai

Figure 7.9 Setting “mA-outputs”

Figure 7.10 Setting “mA-output1 range”