<7. Startup>

7-9

IM 11M12A01-02E

7.9

Setting Display Item

This section briefly describes the display item settings shown in Figure 7.11, “Basic Panel

Display.”

F7.11E.ai

Tag:

21.0

%O

2

17.43mA -Output1

17.43mA -Output2

Primary value

Secondary value

Tertiary value

Figure 7.11 Basic Panel Display

(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.

Then select “Maintenance” in the Execution/Setup display.

(2) Select the “Display setup” from the Maintenance panel display (Figure 7.12).

The Display setup display (Figure 7.13) then appears.

(3) In the above Display setup display, select the “Display item”. The Display item display

(Figure 7.14) then appears. From this display, select the “Primary value” and press the

[Enter] key to display the “Display item” selection display (Figure 7.15).

(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.2, Display Items, enabling the selection of display items in individual display

areas.

Enter

Display setup

Calibration setup

Blow back setup

mA-output loop check

Contact check

Maintenance

F7.12.ai

Enter

Display item

Trend graph

Auto return time: 0min

Language: English

Display setup

F7.13.ai

Figure 7.12 Maintenance Panel Display

Figure 7.13 Display Setup

8th Edition : Jan.13,2012-00