<2. Specifications>

2-12

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

2.4.2

Functions

Display Functions:

Value Display; Displays values of the measured oxygen concentration, etc

Graph Display; Displays trends of measured oxygen concentration

Data Display; Displays various useful data for maintenance, such as cell temperature,

reference junction temperature, maximum/minimum oxygen

concentration, or the like

Status Message; Indicates an alarm or error occurrence with flashing of the

corresponding icon. Indicates status such as warming-up, calibrating,or

the like by icons.

Alarm, Error Display; Displays alarms such as “Abnormal oxygen concentration” or

errors such as “Abnormal cell e.m.f.” when any such status occurs.

Calibration Functions:

Automatic calibration; Requires the Automatic Calibration Unit ZR40H. It calibrates

automatically at specified intervals.

Semi-auto Calibration; Requires the Automatic Calibration Unit. Input calibration direction

on the touchscreen or contact, then it calibrates automatically afterwards.

Manual Calibration; Calibration with opening/closing the valve of calibration gas in

operation interactively with an LCD touchscreen.

Blow back Function:

Output through the contact in the set period and time. Auto/Semi_Auto selectable.

Maintenance Functions:

Can operate updated data settings in daily operation and checking. Display data settings,

calibration data settings, blow back data settings, current output loop check, input/output

contact check.

Setup Functions:

Initial settings suit for the plant conditions when installing the converter.

Equipment settings, current output data settings, alarm data settings, contact data

settings, other settings.

Self-diagnosis:

This function diagnoses conditions of the converter or the detector and indicates when

any abnormal condition occurs.

Password Functions:

Enter your password to operate the analyzer excepting data display.

Individual passwords can be set for maintenance and setup.

Display and setting content:
Measuring related items: Oxygen concentration (vol% O

2

), output current value (mA), air ratio,

moisture quantity (in hot gases) (vol% H

2

O)

Display Items:

Cell temperature (°C), thermocouple reference junction temperature

(°C), maximum/minimum/average oxygen concentration (vol% O

2

), cell

e.m.f. (mV), cell internal resistance (Ω), cell condition (in four grades),

heater on-time rate (%), calibration record (ten times), time (year/month/

day, hour/minute)

Calibration Setting Items: Span gas concentration (vol% O

2

), zero gas concentration (vol% O

2

),

calibration mode (auto, semi-auto, manual), calibration type and

method (zero-span calibration, zero calibration only, span calibration

only), stabilization time (min.sec), calibration time (min.sec), calibration

(day/hour), starting time (year/month/day, hour/minute)