<4. Pipomg>

4-3

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

4.1.2

Connection to the Calibration Gas Inlet

When carrying out calibration, connect the piping (6(O.D) ~4(I.D.) mm tube) from the standard

gas unit to the calibration gas inlet of the detector. First, mount a stop valve (of a quality specified

by YOKOGAWA) on a nipple (found on the local market) as illustrated in Figure 4.2, and mount a

joint (also found on the local market) at the stop valve tip. (The stop valve may be mounted on the

detector prior to shipping the detector.)
Note: Mount the stop valve close to the detector.

F4-2E.ai

Nipple

Stop valve

Tube connection joint

Figure 4.2 Connection to the calibration gas inlet

4.1.3

Connection to the Reference Gas Inlet

• Normally, no piping is required for the reference gas inlet when the equipment uses natural

convection for the reference gas (models ZR22G----C). Leave the plug as it is . If the air

around the detector is polluted and the necessary oxygen concentration (21 vol%O

2

) cannot be

obtained, prepare piping the same as which described in Section 4.2, System 2.
• When the equipment uses instrument air for the reference gas, piping is required as described

in Section 4.2, System 2 (models ZR22G----E or P).

4.1.4

Piping to the High Temperature Probe Adapter

• The sample gas should be at a temperature below 700

°C before reaching the detector

sensor. If the gas is under negative pressure, it should be fed to the detector by suction.
• For usage of the probe adapter when using high temperature detector, refer to Section 3.2.2.
• If the sample gas is under negative pressure, connect the ejector assembly (E7046EC/

E7046EN) as illustrated in Figure 4.3. Mount the pressure gauge as close as possible to the

ejector assembly. However, if the ambient temperature is too high, mount the gauge in a

location with a temperature below 40

°C.

Ejector assembly for
high temperature use

Adapter for
high temperature probe

Detector

F4-3E.ai

Figure 4.3 Mounting the ejector assembly