Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 178

Advertising
background image

<12. Troubleshooting>

12-8

IM 11M12A01-02E

<Locating cause of failure, and countermeasures>

(1) Confirm the following and carry out calibration again: If the items are not within their proper

states, correct them.

a. If the display “Span gas conc” is selected in Calibration setup, the set point should agree

with the concentration of span gas actually used.

b. The calibration gas tubing should be constructed so that the span gas does not leak.

(2) If no alarm is generated as a result of carrying out re-calibration, it is suspected that

improper calibration conditions were the cause of the alarm in the preceding calibration.

In this case, no specific restoration is necessary.

(3) If an alarm is generated again as a result of carrying out re-calibration, deterioration of or

damage to the cell (sensor) is suspected as the cause of the alarm. Replacement of the

cell with a new one is necessary. However, before replacement, carry out the procedure

described in step (3) and later of <Search for cause of failure and taking measure> in

Section 12.2.2.2, Alarm 6: Zero-point Calibration Coefficient Alarm.

12.2.2.4 Alarm 8: EMF Stabilization Time-up Alarm

This alarm is generated if the sensor (cell) voltage has not stabilized even after the calibration

time is up for the reason that the calibration gas (zero gas or span gas) has not filled the sensor

assembly of the detector.
<Cause of alarm>
(1) The flow of the calibration gas is less than normal (a specified flow of 600 ± 60 ml/min).
(2) The length or thickness of the calibration gas tubing has been changed (lengthened or

thickened).

(3) The measuring gas flows toward the tip of the probe.
(4) The sensor (cell) response has deteriorated.

<Locating cause of failure, and countermeasures>
(1) Carry out calibration by passing the calibration gas at the specified flow (600 ± 60 ml/min)

after checking that there is no leakage in the tubing.

(2) If calibration is carried out normally, perform a steady operation without changing the

conditions. If the error occurs again, check whether or not the reason is applicable to the

following and then replace the sensor assembly.

• A lot of dust and the like may be sticking to the tip of the detector probe. If dust is found,clean

the probe (see Section 11.1.1).

In addition, if an error occurs in calibration even after the sensor assembly is replaced, the

influence of sample gas flow may be suspected. Do not let the sample gas flow toward the

tip of the detector probe, for example, by changing the mounting position of the detector.

12.2.2.5 Alarm 10: Cold Junction Temperature Alarm

This alarm is generated when the temperature of the cold junction located at the terminal block of

the detector falls below -25 °C or exceeds 155 °C. Check the following:
Display “C.J.temperature” in the Detailed-data display. If “C.J.temperature” is indicated as 200 °C

or -50 °C, the following can be considered.
(1) Breakage of the cold junction signal wires between the converter and the detector, or the

cable is not securely connected to the connecting terminals.

(2) The positive and negative poles of the cold junction signal wiring are shorted out in the

wiring extension or at the connection terminals.

8th Edition : Jan.13,2012-00

Advertising
This manual is related to the following products: