<8. Detailed Data Setting>

8-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

Setting example 2
If the range minimum (corresponding to 4 mA output) is set to 75 vol%O

2

then range maximum

(corresponding to 20 mA output) must be at least 75x1.3=98 vol%O

2

(rounding decimal part up).

5

15

25

35

45

55

65

75

85

95

5

15

25

35

45

55

65

75

Minimum oxygen concentration, vol%O

2

(for a minimum current of 4 mA)

Minimum-Maximum setting range of oxygen concentration

Maximum oxygen concentration, vol%O

2

(for a maximum current of 20 mA)

Ranges over which oxygen

concentrations can be set

Outside ranges

F8.0E.ai

8.1.3

Setting Output Smoothing Coefficient

When the oxygen concentration in the sample gas fluctuates rapidly, if the measured value is

used for control this can lead to problems with undesirable frequent ON/OFF switching. You can

set a smoothing time constant of between 0 and 255 seconds to reduce the effect. Select the

appropriate output damping constant from the numeric-data entry display. To set 30 seconds,

enter 030.

8.1.4

Selection of Output Mode

You can select whether the relationship between the sample oxygen concentration and the

analog output signal be linear or logarithmic. Press the [Enter] key in the output mode display. A

linear/ logarithmic selection display then appears. Select the desired mode.

NOTE

If you select an output mode of “logarithmic” then regardless of range setting the minimum output

value becomes 0.1 vol%O

2

fixed.

8.1.5

Default Values

When the analyzer is delivered or reset to defaults, the output current default settings by as

shown in Table 8.1.1.

Table 8.1.1 Output Current Default Values

Item

Default setting

Min. oxygen concentration

0 vol%O

2

Max. oxygen concentration 25 vol%O

2

Output damping constant

0 (seconds)

Output mode

Linear