<9. Calibration>

9-7

IM 11M12A01-04E

9.2.1.6

Setting

When setting calibration timing requirements, bear the following precautions in mind:

CAUTION

(1) If the calibration interval is shorter than the sum of hold (output stabilization) time plus calibration

time, the second calibration start time will conflict with the first calibration. In such a case, the

second calibration will not be conducted. (When both zero and span calibrations are to be

performed, the calibration time is double that required for a single (zero or span) calibration.)

(2) For the same reason, if the calibration start time conflicts with manual calibration or semi-

automatic calibration, the current calibration will not be conducted.

(3) If the calibration time conflicts with maintenance service operation, calibration will start after

completing the maintenance service operation (see Section 8.3.1, earlier in this manual).

(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration will be

conducted; a second or later calibration will not be conducted.

(5) If a past date is set to the calibration start day, no calibration will be conducted.

Table 9.1 Parameter Codes for Calibration Setting

Set Item

Parameter

code

Set value

Engineering

unit

Zero gas concentration

B01

Set Zero gas concentration

%O

2

Span gas concentration

B02

Set Span gas concentration

%O

2

Calibration mode

B03

0 Manual calibration
1 Semi-automatic and manual
2 Automatic, semi-automatic, and manual

Hold (Output stabilization) time

B04

0 minutes 0 seconds to 60 minutes 59 seconds MM.SS

Calibration time

B05

0 minutes 0 seconds to 60 minutes 59 seconds MM.SS

Calibration interval

B06

0 days 0 hours to 255 days 23 hours

Date and time

Start date and time

B07

Date and time of first calibration

YY.MM.

DD.HH.MM

Calibration procedure

B08

0 Zero and span
1 Span only
2 Zero only

9.2.1.7

Default Values

When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as

shown in Table 9.2.

Table 9.2

Default Settings for Calibration

Item

Default Setting

Calibration mode

Manual

Calibration procedure

Span - zero

Zero gas (oxygen) concentration

1.00%

Span gas (oxygen) concentration

21.00%

Hold (Output stabilization) time

10 minutes, 00 seconds

Calibration time

10 minutes, 00 seconds

Calibration interval

30 days, 00 hours

Start date and time

00 (YY) 01 (MM) 01(DD) 00:00