<4. Piping>

4-7

IM 11M12A01-04E

4. Set the instrument air pressure higher than furnace internal pressure.

5. Completely open the stop valve in front of the reference gas outlet and, after turning on

instrument air flow, start furnace operation. As furnace internal pressure rises, confirm that

instrument air continues to flow and adjust the valve or increase supply pressure if necessary.

6. After furnace internal pressure stabilizes, adjust flow.

7. If furnace operation is stopped, stop instrument air flow and completely shut the stop valve in

front of the reference gas outlet. You may leave reference gas flowing if you wish.

CAUTION

• Use suitable cable glands to completely seal the detector. As far as possible do not stop the

instrument air flow, to prevent the sample gas from entering the detector and damaging the

zirconia cell.

• Connect the stop valve, which is at the calibration gas inlet, directly to the equipment.

If piping connections are made between the detector and the needle valve, condensation

will result inside the piping and cause the sensor to be damaged when the calibration gas is

introduced.

Figure 4.13 illustrates an example of System 2 using the analyzer with pressure compensation.
Supplying the air pressure (flow) may vary depending on the furnace pressure. It is recommended to

use a flow gauge and an air set meeting the furnace pressure.

~

ZR202G Integrated type Zirconia Oxygen Analyzer

with pressure compensation

F1.2E.ai

ZA8F flow setting unit

Reference gas

Calibration gas

Needle

valve

Flowmeter

Instrument air

Air Set

Pressure

regulator

Zero gas cylinder

Calibration gas unit case

Stop valve

or

Check valve

Span gas(Same as Zero gas calibration)

100 to 240 V AC

Contact input
Analog output, contact output

Digital output (HART)

Figure 4.13 Illustrates an example of System 2 using the analyzer with pressure

compensation.

CAUTION

• When using the ZA8F Flow Setting Unit and the ZR20H Automatic Calibration Unit, please note that

the supplying airflow (pressure) will vary depending on the furnace pressure.