5.4 Wiring for Contact Output

The equipment can output a maximum of two contact signals. These contact outputs can be used for

different applications such as a low alarm or high alarm.
Do the contact output wiring according to the following requirements.

DO-1
DO-1

DO-2
DO-2

F5.6E.ai

#2 Output

Analyzer

Terminal box

Annunciator or the like

#1 Output

Figure 5.6

Contact Output Wiring

5.4.1 Cable Specifications

The number of cores varies depending on the number of contacts used.

5.4.2 Wiring Procedure

(1) M4 screws are used for the terminals. Use crimp-on terminals appropriate for M4 terminal

screws for cable connections.

(2) The contact output relays are rated 30 V DC 3A, 250 V AC 3A. Connect a load (e.g. pilot lamp

and annunciator) within these limits.

5.5 Wiring for Contact Input

The converter can execute specified function when receiving contact signals.
To use these contact signals, proceed wiring as follows:

Converter

Terminal box

DI-1

DI-C

DI-2

Contact input 1

Contact input 2

F5.7E.ai

Figure 5.14

Contact Input Wiring

5.5.1 Cable Specifications

Use a 2-core or 3-core cable for this wiring. Depending on the number of input(s), determine which

cable to use.