<10. Other Functions>

10-14

IM 11M12A01-04E

Contact-related Items in Group E

Code

Item

Tuning

Engineering unit

Default setting

E01 Selection of input contact 1 0 Invalid

Invalid

1 Calibration gas pressure decrease
2 Measurement range change
3 Calibration start
4 Detection of non-combusted gas

E02 Selection of input contact 2 0 Invalid

Invalid

1 Calibration gas pressure decrease
2 Measurement range change
3 Calibration start
4 Detection of non-combusted gas

E03 Selecting action of input

contact 1

0 Action with closed contact

Action with closed

contact

1 Action with open contact

E04 Selecting action of input

contact 2

0 Action with closed contact

Action with closed

contact

1 Action with open contact

E10 Selecting action of output

contact 1

0 Action with closed contact

(normally energized)

Action with closed

contact

1 Action with open contact (normally

de-energized)

E20 Output contact 1 error

0 No action

No action

1 Action

E21 Output contact 1, high-high

alarm

0 No action

No action

1 Action

E22 Output contact 1, high

alarm

0 No action

No action

1 Action

E23 Output contact 1, low alarm 0 No action

No action

1 Action

E24 Output contact 1, low-low

alarm

0 No action

No action

1 Action

E25 Output contact 1, during

maintenance

0 No action

Action

1 Action

E26 Output contact 1, during

calibration

0 No action

No action

1 Action

E27 Output contact 1,

measurement range

change

0 No action

No action

1 Action

E28 Output contact 1, during

warm up

0 No action

Action

1 Action

E29 Output contact 1,

calibration gas

pressure decrease

0 No action

No action

1 Action

E32 Output contact 1, detection

of non- combusted gas

0 No action

No action

1 Action