3 entering parameter code selection display, Entering parameter code selection display -5, Passno – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 70: Caution

Advertising
background image

<7. Startup>

7-5

IM 11M12A01-04E

Table 7.2

Display Functions

Display

Function and item to be set

Basic panel

Displays the oxygen concentration in normal operation, or displays the detector

heater temperature while warming up. If an error or alarm arises,

the corresponding error or alarm number appears.

Password entry

Enters the password for the parameter code selection display.

Group A setup

Displays detailed data, such as the cell voltage or temperature.

Group B setup

Sets and performs calibration.

Group C setup

Sets analog output.

Group D setup

Sets an alarm.

Group E setup

Sets the input and output contacts.

Group F setup

Selects the type of equipment and sets the parameters for computation.

Group G setup

Performs the current-loop or contact checks.

7.4.3 Entering Parameter Code Selection Display

This section briefly describes the password entry procedure for entering the parameter code selection

display. The password is 1102 - it cannot be changed to a different password.

Switch operation

Display

Description

>

ENT

21.0%

Warm-up is complete, and the basic panel is now displayed.

>

ENT

PASSno

Continuously touch the [ENT] key for at least three seconds
to display "PASSno."

>

ENT

0

000

Touch the [ENT] key again. This allows you to change the leftmost
digit that is flashing.

>

ENT

1

000

Set the password 1102. If you touch the [∧] key, the digit that is flashing will
be 1.

>

ENT

1

0

00

Touch the [>] key to move the position of the digit that is flashing to the
right one digit.

>

ENT

1

1

00

Touch the [∧] key to change the numeric value to 1.

>

ENT

110

0

Touch the [>] key again to move the position of the digit that is flashing to
the right one more digit. Continuously touch the [>] key, and the position of
the digit that is flashing will move continuously to the right.

>

ENT

110

2

Touch the [∧] key to change the numeric value to 2. Continuously touch [∧]
key, and the numeric value increases continuously.

>

ENT

1102

If you touch the [ENT] key, all the digits flash.

>

ENT

A

01

Touch the [ENT] key again to display A01 on the parameter code selection
display.

The symbol [

] indicates that the key is being touched. Light characters indicate that the digits are flashing.

CAUTION

• If no key is touched for at least 20 seconds during password entry, the current display will

automatically switch to the basic panel display.

• If no key is touched for at least 10 minutes during parameter code selection, the current display

will automatically switch to the basic panel display.