4.2 Piping for System 2

Piping in System 2 is illustrated in Figure 4.7.

~

ZR202G Integrated type Zirconia Oxygen Analyzer

F1.2E.ai

ZA8F flow setting unit

Reference gas

Calibration gas

Needle

valve

Flowmeter

Instrument air

Air Set

Pressure

regulator

Zero gas cylinder

Calibration gas unit case

Stop valve

or

Check valve

Span gas(Same as Zero gas calibration)

100 to 240 V AC

Contact input
Analog output, contact output

Digital output (HART)

Figure 4.7

Piping for System 2

System 2 illustrated in Figure 4.7 requires piping as follows:
• Mount the check valve or the stop valve through a nipple to the calibration gas inlet of the equipment.

4.2.1 Piping Parts for System 2

Check that the parts listed in Table 4.2 are provided.
Table 4.2

Piping Parts

Equipment

Piping location

Parts Name

Description

Oxygen Analyzer Calibration gas inlet Stop valve or check valve Stop valve (L9852CB or G7016XH

recommended by YOKOGAWA

Check valve (K9292DN or K9292DS)

provided by YOKOGAWA

Nipple *

Rc1/4 or 1/4 NPT

General parts

Zero gas cylinder

User' s scope

Pressure Regulator

(G7013XF or G7014XF) recommended by

YOKOGAWA

Joint for tube connection Rc1/4 or 1/4 NPT

General parts

Reference gas inlet Air set

(G7003XF/ K9473XK or G7004XF

/ K9473XG) recommended by YOKOGAWA

Joint for tube connection Rc1/4 or 1/4 NPT

General parts

Note: Parts with marking * are used when required.

General parts can be found on the local market.

4.2.2 Piping for the Calibration Gas

This piping is to be installed between the zero gas cylinder and the ZA8F flow setting unit, and

between the ZA8F flow setting unit and the ZR202G analyzer.
The cylinder should be placed in a case assembly E7044KF or the like to avoid any direct sunlight or

radiant heat so that the gas cylinder temperature may not exceed 40°C.
Mount the pressure regulator (recommended by YOKOGAWA) on the cylinder.
Mount the stop valve or the check valve (recommended by YOKOGAWA) through the nipple (found

on the local market) at the calibration gas inlet of the equipment as illustrated in Figure 4.8. (The stop

valve or the check valve may have been mounted on the equipment when shipped.) Connect the

flow setting unit and the analyzer to a 6mm (O.D.) x 4mm (I.D.) (or nominal size 1/4 inches) or larger

stainless steel pipe.