Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 132

Advertising
background image

<11. Inspection and Maintenance>

11-3

IM 11M12A01-04E

(4) Use tweezers to pull the contact out of the groove.

(5) Clean the sensor assembly, especially the metal O-ring contact surface to remove any

contaminants adhering to that part. If you can use any of the parts from among those removed,

also clean them up to remove any contaminants adhering to them.

(Once the metal O-ring has been tightened, it can no longer be used. So, be sure to replace it.)

3. Part assembly procedure

(1) First, install the contact. Being careful not to cause irregularities in the pitch of the coil spirals

(i.e., not to bend the coil out of shape), place it in the ringed groove properly so that it forms a

solid contact.

Groove in which the contact

(E7042BS) is placed

F11.2E.ai

Figure 11.2

Installing the Contact

(2) Next, make sure that the O-ring groove on the flange surface of the sensor (cell) is clean. Install

the metal O-ring in that O-ring groove, and then insert the sensor (cell) in the probe while turning

it clockwise. After inserting it until the metal O-ring comes in contact with the probe’s O-ring

contact surface, properly align the U-shaped-pipe insertion holes with the bolt openings.

(3) Attach the U-shaped pipe to its support, then fully insert the U-shaped pipe, filter and its support

into the probe.

(4) Coat the threads of the four bolts with antiseize grease and then screw them in along with

the washers. First, tighten the four bolts uniformly by hand, and then use a torque wrench to

tighten all areas of the metal O-ring uniformly, that is, to make sure the sensor flange is perfectly

horizontal to the O-ring’s working face in the probe. This is done by tightening first one bolt and

then its opposing bolt each 1/8 turn, and then one of the other bolts followed by its opposing bolt,

each also 1/8 turn.

This continues in rotating fashion until they are all fully tightened with the torque wrench preset

to approximately 5.9 Nm. If they are not uniformly tightened, the sensor or heater may be

damaged. Check with light that there is no gap between sensor flange and probe.

Replacement of the sensor assembly is now complete. Install the detector and restart operation.

Calibrate the instrument before making a measurement.

Advertising