4 setting oxygen concentration alarms, 1 alarm values, 2 alarm output actions – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 89: Setting oxygen concentration alarms -6 8.4.1, Alarm values -6, Alarm output actions -6

Advertising
background image

<8. Detailed Data Setting>

8-6

IM 11M12A01-04E

8.4 Setting Oxygen Concentration Alarms

The analyzer enables the setting of four alarms high-high, high, low, and low-low alarms depending

upon the oxygen concentration. The following section sets forth the alarm operations and setting

procedures.

8.4.1 Alarm Values

(1) High-high and high alarm values

High-high alarms and high alarms are issued when they are set to be detected with parameter

codes

「D41」 and 「D42」, and if the measured values exceed the preset oxygen concentration

values specified with

「D01」 and 「D02」.

The oxygen alarm set values can range from 0 to 100% O

2

.

(2) Low and low-low alarm values

Low alarms and low-low alarms are issued when they are set to be detected with parameter

codes

「D43」 and 「D44」, and if the measured values are lower than the preset oxygen

concentration values specified with

「D03」 and 「D04」.

The oxygen alarm set values can range from 0 to 100% O

2

.

8.4.2 Alarm Output Actions

If the measured values of the oxygen concentration fluctuate between normal (steady-state) values

and alarm setting, there may be a lot of alarm-output issuing and canceling. To avoid this, set the

delay time and allow for hysteresis for alarm canceling under the alarm output conditions, as Figure

8.2 shows. When the delay time is set, an alarm will not be issued so quickly even if the measured

value differs from the steady-state and enters the alarm setpoint range. If the measured value remains

within the alarm setpoint range for a certain period of time (for the preset delay time), an alarm will

result. On the other hand, there will be a similar delay each time the measured value returns to the

steady state from the alarm setpoint range (canceling the alarm status). If hysteresis is set, alarms will

be canceled when the measured value is less than or more than the preset hysteresis values. If both

the delay time and hysteresis are set, an alarm will be issued if the measured value is in the alarm

setpoint range and the delay time has elapsed. When the alarm is reset (canceled), it is required that

the measured value be beyond the preset hysteresis value and that the preset delay time. Refer to

Figure 8.2 for any further alarm output actions. The delay time and hysteresis settings are common to

all alarm points.

Delayed time:

5 seconds

7.5%

High alarm setpoint

5.5%

Oxygen concentration

Alarm output ON

OFF

Hysteresis

2.0%

A

B

C

D

Alarm range

Delayed time:

5 seconds

Delayed time:

5 seconds

F8.2E.ai

Figure 8.2

Alarm Output Action

Advertising