<7. Startup>

7-10

IM 11M12A01-04E

Table 7.5

Minimum and Maximum Value Setting Procedure

Switch operation

Display

Description

>

ENT

A

01

Display after the password has been entered.

>

ENT

C

01

Set the oxygen concentration at 4 mA. Change the parameter code to C11.
Touch the [∧] key to switch to Group C.

>

ENT

C

0

1

Touch the [>] key to move the position of the digit that is flashing to the right.

>

ENT

C

1

1

Touch the [∧] key to enter the numeric value 1.

>

ENT

0

00

Touch the [ENT] key to display the current set value (0% O

2

has been set).

>

ENT

0

0

0

Touch the [>] key to move the position of the digit that is flashing to the right.

>

ENT

0

1

0

Touch the [∧] key to change the numeric value to 1.

>

ENT

010

If you touch the [ENT] key, all the digits flash.

>

ENT

010

Touch the [ENT] key again to stop the flashing.

>

ENT

C

11

Touch the [ENT] key once again, and the display will switch to the parameter code
selection display.

>

ENT

C1

1

Set the oxygen concentration at 20 mA. Touch the [>] key to move the position of
the digit that is flashing to the right.

>

ENT

C1

2

Touch the [∧] key to enter the numeric value 2.

>

ENT

0

25

Touch the [ENT] key to display the current set value.

>

ENT

025

Touch the [>] key to move the position of the digit that is flashing to the right.

>

ENT

02

0

Touch the [∧] key to change the numeric value to 0. The numeric value will change
from 5 to 6 ... to 9 and back to 0.

>

ENT

020

If you touch the [ENT] key, all the digits flash.

>

ENT

020

Touch the [ENT] key again to stop the flashing.

>

ENT

C

12

Touch the [ENT] key once again to switch to the parameter code selection display.

>

ENT

Basic

panel

display

Touch the [>] key together with the [ENT] key to return to the basic panel display.
(This is not required if you proceed to make another setting.)(The displayed
numeric characters indicate the measurement gas concentration.)