C H A P T E R 2 - I N S T A L L A T I O N

15

Figure 16: Correcting a Field Mill using a Reference Field Mill


permanent

reference

correction


Once you have calculated the required sensitivity reduction (correction factor) you can

calculate the resistor value needed for the Sensitivity Plug.


k

correction

k

resistor

100

)

1

(

100


For most installations it is not necessary to correct using a reference field mill. It is usually

sufficient to choose a sensitivity plug which produces a fair-weather electric field of

approximately 0.1 kV/m and does not saturate (above -20 kV/m or +20kV/m) during
most thunderstorms. Electric field trends and sudden changes (lightning) are of much

more interest than an exact absolute value. Alarm set points can be lowered to increase

alarm sensitivity or increased to reduce the frequency of nuisance alarms.