C H A P T E R 2 - I N S T A L L A T I O N

16

Grounding Detail

The EFM-100 field mill needs to be grounded for proper operation and for lightning and

electrical safety. Connect the connector end of the green ground wire to the field mill

ground connector. Connect the other end of the green ground wire and the field mill
mounting hardware to your building’s central ground through a #6 bare ground wire or

other suitable wire according to your local electrical code. Grounding wire and clamps are

available at your local hardware store or electric supplier.


IMPORTANT

The EFM-100 field mill needs to be grounded for proper operation and for

lightning and electrical safety. Connect the field mill green ground wire

and the field mill mounting hardware to your building central ground

through a suitable sized grounding wire according to your local electrical

code.