Calibrating Gauges

18

FastCalXP Operation Manual

To Change the Number of Exercise Cycles Associated with a Profi le

1

choose

Manage Database

from the Tools menu and click on the

Profi les button

.

2

Select the profi le of interest and click on the

Edit button

.

3

Enter the desired number of exercise cycles in the

Number of exercise cycles

text box.

Matching the Target

Two diff erent approaches can be taken while stepping through a calibration profi le: matching the gauge to the Target Pressure or matching the reference to

the Target Pressure.

During a calibration, the

Target Pressure

is the next profi le value and the

Reference

and

Gauge

pressures are indicated below. The Gauge value is, by default, the

same value as the Target Pressure. However, the value can be changed by the user to match what is read from the gauge.

To Match the Gauge to the Target Pressure

1

Read the gauge while turning the handle of the GaugeCalHP.

2

Stop when the gauge reaches the Target Pressure value.

Matching the gauge allows for much faster calibration since interpolation is not required to read the gauge and the Gauge value does not have to be

entered.