Crystal FastCalXP Calibration Software User Manual

Page 39

Advertising
background image

Managing the Database

37

FastCalXP Operation Manual

Use Custom Certifi cate Template

When

Use custom certifi cate

template is checked, you can override the default Excel template defi ned in the Options dialog. This allows diff erent manufactur-

ers or models to have unique certifi cate formats.

Cal Interval

Enter the

Cal interval

(in days) which will be used for calibration scheduling.

Transmitter

Select

Transmitter

when you will select which parameters of the transmitter to measure during calibration.

Measure mA

Select

Measure mA

when calibrating the analog output of the transmitter. Specify the Low/High pressure range as well as the corresponding Low/High mA

output of the transmitter.

Measure HART

If the model is a smart transmitter, select

Measure HART

to view digital output of the transmitter. If the transmitter has an external display check

Measure

Displayed

to include the external display in the certifi cate. You can select either

Measure Displayed

or

Measure HART

but not both for a model.

Accuracy

There are three ways to defi ne the accuracy of the model. Many gauges will have an ASME or ISO/IEC grade which can be selected from the pull-down list. If

the accuracy is not listed, you may be able to simply enter a

% of span

accuracy.

Alternatively, select

Custom

to defi ne a custom accuracy by specifying a series of pressure ranges and accuracies associated with those ranges. Click the

Add

button

to include a range and accuracy in the specifi cation. When the accuracy is fully specifi ed, click

OK

.