Head correction calculator – Crystal FastCalXP Calibration Software User Manual

Page 34

Advertising
background image

Settings and Options

32

FastCalXP Operation Manual





HEAD CORRECTION CALCULATOR

The Head Correction Calculator is accessed via a button under the Calibration Settings tab.

The calculator uses

Fluid density

,

Gravity

and

Height change

to calculate a correction that is added to the pressure measured by the reference.

Four standard fl uid densities, for

Isopropyl Alcohol at 25°C

,

Pure Water at 4°C

,

Pure Water at 20°C

and

Air at 15°C

and sea level, can be applied using the buttons

in the Head Correction box.

The local gravity can be estimated using the local

Latitude

and the altitude above

Mean sea level

.

Enter these values and click on the

Calculate Gravity button

.

The calculated gravity will automatically be entered into the

Gravity

fi eld in the Head Correction box.

When the appropriate parameters are entered, click on the

Calculate Head Correction button

to display the calculated Head Correction. Click

OK

to insert the

correction into the head correction box in the Calibration Settings.