Crystal FastCalXP Calibration Software User Manual

Page 30

Advertising
background image

Settings and Options

28

FastCalXP Operation Manual

Save Certifi cates To

The calibration certifi cates created by FastCalXP will be saved in the directory specifi ed in

Save certifi cates to.

This directory is set the fi rst time the software is

run after installation or when a new database is created and can be changed here using the Browse button.

Certifi cate Naming Template

The

Certifi cate naming template

is used by FastCalXP to name the calibration certifi cates. In general, certifi cates are numbered with the

Next certifi cate

number

incrementing automatically each time a certifi cate is created. Use number signs (#’s) in the Certifi cate naming template to defi ne the number of digits

to be used.

Calibration Representative and Quality Representative

The

Calibration representative

and the

Quality representative

must be listed users and can be selected using the pull-down lists. In the certifi cate template

provided with the software these names will be listed, with signature lines, on the fi rst page of each certifi cate.

Calibration Settings

The Calibration Settings defi ne the default Calibration Method.

Actual

Uses the pressure reading from the reference at the moment the Next Point button is clicked.

Peak

Automatically advances to the Next Point when a peak is detected.

Head Correction

If the gauge and the reference do not sit at the same height, a

Head correction

can be applied. Enter a correction to be added to the reference pressure, or use

the Correction calculator to calculate a correction given the specifi c conditions of the calibration. Refer to the discussion of the

Head Correction Calculator

on page 32

.