Take the as received data, Take the as left data, Generate and print the certifi cate – Crystal FastCalXP Calibration Software User Manual

Page 9

Advertising
background image

Quick Start

7

FastCalXP Operation Manual





TAKE THE AS RECEIVED DATA

Perform a New Calibration.

1

Click the

Perform As Received Cal button

.

2

The Calibration Wizard will auto-detect the reference as described in the discussion of

References.

If a new reference is detected, the Calibration Wizard will prompt you to enter the current certifi cate information for the reference. If using an nVision with

two modules you will be prompted to select one of the modules to use for the calibration.

3

If the profi le requires it, exercise the gauges by increasing and decreasing the pressure over the full range of the gauge one or more times. Refer to the

GaugeCalHP Operation Manual

or the discussion of

Exercising on page 17

for more information.

The Calibration Window will then open. Refer to description of the

Calibration Window

in the

Menus and Windows

section

.

4

Select

Actual

or

Peak

mode and step through the calibration. Refer to the discussions of

Calibration Strategies

,

Actual Mode Calibration

, and

Peak Mode

Calibration

later in the manual.

Load Existing As Received Data.

If the As Received calibration has been completed in a previous session, complete the As Received Step by clicking the

Load Previous Data button

.





TAKE THE AS LEFT DATA

If the gauge failed the As Received calibration and adjustments have been made to the gauge, click on the

Perform As Left Cal button.

Follow steps 2 through

4 in the

Perform a New Calibration

instructions above.

Alternatively, if the gauge passed the As Received calibration and no adjustments have been made to the gauge, FastCalXP can use the As Received calibra-

tion for the As Left calibration. Click on the

Use As Received for As Left button

.





GENERATE AND PRINT THE CERTIFICATE

1

Enter any notes regarding the calibration or modifi cations made to the gauge in the Calibration Notes / Modifi cations box and click on the

Save Data button

in the Calibration Status box. Notes cannot be added after the certifi cate has been generated.

2

Create the calibration certifi cate by clicking on the

Generate Certifi cate button

.

FastCalXP will create the certifi cate and open the certifi cate by running

Excel. Go to the Excel window with the current certifi cate and print the certifi cate.

Refer to the

Creating Certifi cates

section for more information on working with certifi cates.

Advertising
This manual is related to the following products: