4 statistical calculation mode, 4-1 getting started – A&D FX-i SERIES User Manual

Page 18

Advertising
background image

16

5-4 Statistical Calculation Mode

The statistical calculation mode statistically calculates the weight data, and displays or outputs the

results. To use the statistical calculation mode, set the "Application function (apf)" parameter of

“Application (ap fnc)” in the function table to “2”, as described below.

Statistical items available are number of data, sum, maximum, minimum, range

(maximum-minimum), average, standard deviation and coefficient of variation. What statistical items

to output can be selected from the three modes in the function table.

The wrong data input can be canceled by the key operation, if immediately after the input.
Turning the balance off will delete the statistical data.
The standard deviation and coefficient of variation are obtained by the equation below:

Standard deviation=

Standard deviation

Coefficient of variation (CV)=

Average

5-4-1 Getting Started

Switching to the Statistical Function Mode (Changing The Function Table)

1 Press and hold the SAMPLE key until ba5fnc of

the function table is displayed, then release the key.



2 Press the SAMPLE key several times to display

1

ap fnc .



3 Press the PRINT key to display apf 0 .



4 Press the RE-ZERO key to display apf 2 .

To select statistical items to output, go to step 5.

To store the statistical function mode setting, go to

step 7.

To disable the statistical calculation mode, press the

RE-ZERO key to select apf 0 .

where Xi is the i-th weight data,

N is number of data.

x 100 (%)