A&D FX-i SERIES User Manual

Page 29

Advertising
background image

27

5 Confirm that there is nothing on the pan and press the

PRINT key. The balance measures the zero point and

displays the measured value. Do not allow vibration or

drafts to affect the balance.

The balance displays the target weight value.

6 Place the displayed target weight on the pan and press

the PRINT key. The balance measures the target weight

and displays the measured value. Do not allow vibration

or drafts to affect the balance.

7 The balance displays end . Remove the weight from the

pan.

8 When the GLP output is set, the balance displays glp

1

and outputs “Calibration Test Report”.

9 The balance will automatically return to the weighing

mode.