7-2 Calibration Using an External Weight

This function calibrates the balance using an external weight.

Operation

1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until Calout is

displayed, and then release the key.

3 The

balance

displays Cal 0 .

If you want to change the calibration weight (a

list of usable weights is shown on page 23),

press the SAMPLE key and proceed to step 4.

If you use the calibration weight value stored in

the balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE key

To switch the display condition

to: “All of the segments

blinking” (calibration weight

selection mode) or “The last

two digits blinking” (value

adjustment mode).

RE-ZERO key (To increase the value)

MODE key (To decrease the value)

To select the calibration

weight or adjust the value.

PRINT

key

To store the new weight

value. Even if the AC

adapter is removed, the data

is maintained in non-volatile

memory.

CAL key

To cancel the operation and

return to Cal 0 .