9-4 Description of the Class “Environment, Display”

Condition (

Cond

)

Cond 0

This parameter is for sensitive response to the fluctuation of a weight value.

Used for powder target weighing, weighing a very light sample or when

quick response weighing is required.

After setting, the balance displays FAST.

Cond 2

This parameter is for stable weighing with slow response. Used to prevent a

weight value from drifting due to vibration or drafts.

After setting, the balance displays SLOW.

Note
With “Hold function (

Hold)” set to “ON (1)”, this item is used to set the

averaging time.

Stability band width (

5t-b )

This item controls the width to regard a weight value as a stable value. When the fluctuation per

second is less than the parameter, the balance displays the stabilization indicator and outputs the

data. The parameter influences the “Auto print mode”

5t-b 0

This parameter is for sensitive response of the stabilization indicator. Used

for exact weighing.

5t-b 2

This parameter ignores slight fluctuation of a weight value. Used to prevent

a weight value from drifting due to vibration or drafts.

Note
With “Hold function (

Hold)” set to “ON (1)”, this item is used to set the

stabilization range.

Hold function (

Hold ) (Animal weighing mode)

This function is used to weigh a moving object such as an animal.

When the weight data is over the weighing range from zero and the display fluctuation is within the

stabilization range for a fixed period of averaging time, the processing indicator illuminates and the

balance displays the average weight of the animal. When the animal is removed from the weighing

pan, the display returns to zero automatically.

This function is available only when the hold function parameter is set to “

1” (the animal mode

indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected.

The stabilization range and averaging time are set in “Condition (

Cond)” and “Stability band width (5t-b)”.

Weighing range

Averaging time

Stabilization range

FX-120

i /200 i / 300 i

0.200 g or over

Cond 0

5t-b 0

FX-1200

i /2000 i / 3000 i 2.00 g or over Cond 1

5t-b 1

Cond 2

2 seconds Faster
4 seconds |
8 seconds More accurate

5t-b 2

Lesser

|

Greater