2 acknowledge code and error codes – A&D FX-i SERIES User Manual

Page 54

Advertising
background image

52

15-2 Acknowledge Code and Error Codes

When the “AK, Error code (

erCd)” parameter of “Serial interface ( 5if )” is set to “1”, the balance

outputs <AK> code or an error code for each command as follows:

<AK> (06h) Acknowledge in ASCII code.

When the balance receives a command to request data and can not process it, the balance

transmits an error code (EC, Exx).

When the balance receives a command to request data and can process it, the balance

outputs the data.

When the balance receives a command to control the balance and can not process it, the

balance transmits an error code (EC, Exx).

When the balance receives a command to control the balance and can process it, the balance

transmits the acknowledge code.

Among commands to control the balance, the following transmit the acknowledge code both

when the balance receives the command and when the balance has accomplished the
command. If the command can not be processed properly, the balance transmits an error code
(EC, Exx). This error can be released using the CAL command.

CAL command (Calibration command)

ON command (Display ON command)

P command (Display ON/OFF command)

R command (RE-ZERO command)

R command example

Advertising