3. PRECAUTIONS

To get the optimum performance from the balance and acquire accurate weight data, note the

following:

3-1 Before Use

Install the balance in an environment where the

temperature and humidity are not excessive. The best

operating temperature is about 20°C / 68°F at about

50% relative humidity.

Install the balance where it is not exposed to direct

sunlight and it is not affected by heaters or air

conditioners.

Install the balance where it is free of dust.

Install the balance away from equipment which

produces magnetic fields.

Install the balance in a stable place avoiding vibration

and shock. Corners of rooms on the first floor are best,

as they are less prone to vibration.

The weighing table should be solid and free from

vibration, drafts and as level as possible.

Level the balance by adjusting the leveling feet and

confirm it using the bubble spirit level.

Ensure a stable power source when using the AC

adapter.

Plug in the AC adapter and warm up the balance for at

least 30 minutes.

Calibrate the balance before use or after having

moved it to another location.

Caution

Do not install the balance where flammable or corrosive gas is present.

3-2 During Use

Discharge static electricity from the material to be

weighed (hereinafter referred to as the sample). When

a sample could have a static charge, the weight data is

influenced. Ground the balance and try the following:

Eliminate static electricity using the optional
AD-1683 DC static eliminator.

Try to keep the ambient humidity above 45%RH.
Use a metal shield case for a charged sample.
Wipe charged plastic samples with a moistened

cloth.