A&D FX-i SERIES User Manual

Page 34

Advertising
background image

32

Class

Item and Parameter

Description

0

Key mode

Accepts the PRINT key only when
the display is stable.

1

Auto print mode A
(Reference = zero)

2

Auto print mode B
(Reference = last stable value)

Outputs data when the display is
stable and conditions of

ap-p, ap-b

and the reference value are met.

3

Stream mode

Outputs data continuously.

4

Key mode B (Immediately)

Accepts the PRINT key regardless
of the display condition.

5

Key mode C (When stable)

Accepts the PRINT key immediately
when the display is stable, or waits
for the display to be stable when not.

prt

Data output mode

6

Interval output mode

Uses interval output mode.

0

Plus only

Displayed value>Reference

1

Minus only

Displayed value<Reference

ap-p

Auto print polarity

2

Both

Regardless of displayed value

0

10 digits

1

100 digits

ap-b

Auto print difference

2

1000 digits

Difference between reference value
and displayed value

0

Every measurement

1

2 seconds

2

5 seconds

3

10 seconds

4

30 seconds

5

1 minute

6

2 minute

7

5 minute

int

Interval time

8

10 minute

Interval time for the interval output
mode
(With

prt 6)

0

No pause

pU5e

Data output pause

1

Pause (1.6 seconds)

Selects the data output interval.

0

Not used

at-f

Auto feed

1

Used

Selects whether or not automatic
feed is performed.

0

No output

1

AD-8121 format

info

GLP output

2

General data format

Selects GLP output method.

0

Not used

dout

Data output

ar-d

Zero after output

1

Used

Adjusts zero automatically after data
is output

Factory settings. Digit is a unit of minimum weighing value