5 description of the item “data output mode – A&D FX-i SERIES User Manual

Page 38

Advertising
background image

36

9-5 Description of the Item “Data Output Mode”

The parameter setting of the “Data output mode (

prt )” applies to the performance when the data is

transmitted using the RS-232C serial interface.

Key mode

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the

weight data and the display blinks one time.

Required setting

dout

prt 0 Key

mode

Auto print modes A and B

When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference”

and reference value are met, the balance outputs the weight data.

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the

data and the display blinks one time.

Mode A: Required setting

dout prt

1

Auto print mode A (reference = zero)

dout ap-p

Auto print polarity

dout ap-b

Auto print difference

Example

“For outputting the weight value each time a sample is added, with

ar-d” set to “1” (to adjust zero after the data is output).”

Mode B: Required setting

dout prt

2

Auto print mode B (

reference = last stable value)

dout ap-p

Auto print polarity

dout ap-b

Auto print difference

Example

“For outputting the weight value while a sample is added.”

Stream mode

The balance outputs the weight data continuously regardless of the display condition. The display

does not blink in this mode.

Required setting

dout prt

3 Stream

mode

ba5fnc 5pd

Display refresh rate

5if bp5 Baud

rate

Example

“For monitoring data on a computer”

Caution

The balance may not transmit the data completely at the specified refresh rate and baud rate.
Set the baud rate higher.

Key mode B

When the PRINT key is pressed, the balance outputs the weight data, regardless of the display

condition. The display does not blink in this mode.

Key mode C

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the

weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on

and outputs the data. The balance blinks one time in this mode.

Advertising