Boonton 9240 series rf voltmeter, Getting started 3-13 – Boonton 9240 RF Voltmeter User Manual

Page 39

Advertising
background image

Boonton 9240 Series RF Voltmeter


RS232 Menu. The RS232 menu is used to configure the Model 9240 for serial communications over the RS-232 bus. An
example of the submenu is shown in Figure 3-10 and an explanation of the commands, parameters and factory defaults is
given in Table 3-5.

R T N

B A U D

R A T E

> 3 8 4 0 0

D A T A

B I T S

8

S T O P

B I T S

1

P A R I

T Y

N O N E

E O S

L S T N

L F

E O S

T A L K E R C R L F

S R Q M A S K

0

Figure 3-10. RS232 Menu Display


Table 3-5 RS232 MENU Functions


Function Description

Parameters

Defaults


RTN

Returns the instrument to the

n/a

previous

menu.


BAUD RATE

Rate at which data is

300, 600, 1200, 2400, 4800,

38400

transferred over the bus.

9600, 19200, 38400, 57600,

115200

DATA BITS

Number of data bits in a message.

7, 8

8


STOP BITS

Number of stop bits in a message.

1, 2

1


PARITY

Parity bit mode in a message.

ODD, EVEN, NONE

NONE


EOS LSTN

End of string indicator for

LF, CR, CRLF, NONE

LF

received

messages.

Where:

LF = Line Feed

CR = Carriage Return

CRLF = Carriage Return

and Line Feed


EOS TALKER

End of string character sent

LF, CR, CRLF, NONE

CRLF

with transmitted messages.


SRQ MASK

Service request interrupt mask.

0 to 255

0

See Table 4-7 for bit

Where:

descriptions.

255 enables all interrupts

Getting Started

3-13

Advertising