Boonton 9240 Series RF Voltmeter

5.5.2 CALCulate Subsystem

The CALCulate group of the command subsystem is used to configure post acquisition data processing. Functions in the
CALCulate subsystem are used to configure the measurement mode and control which portions of the acquired measurement
data is used and how it is processed to yield a finished measurement. In addition to measurement mode, CALCulate is used
to define mathematical operations, measurement units, and limit monitoring. The numeric suffix of the CALCulate program
mnemonic in the CALCulate commands refers to a processing and display “channel”, that is CALCulate1 and CALCulate2
represent the voltmeter’s Channel 1 and Channel 2 functions. The CALCulate commands generally DO NOT affect the data
acquisition portion of the measurement (see the SENSe subsystem, Section 5.5.11 ). In a signal-flow block diagram, the
CALCulate block operations will follow those of the SENSe block. Note that CALCulate2 commands will generate an error
if used with a single channel Model 9241.

CALCulate:LIMit:CLEar[:IMMediate]

Description:

Clear all latched alarms for the selected channel.

Syntax: CALCulate[1|2]:CLEar[:IMMediate]

Argument: None

CALCulate:LIMit:FAIL?

Description:

Returns the status of all limit alarms for the specified channel in five flags. The first flag is the

logical sum of the remaining four. They are: low-limit active, high-limit active, low-limit latched

and

high-limit

latched.

Active means that the limit is exceeded when the command is executed;

Latched means that the limit has been exceeded since the last limit clear command.


Syntax:

CALCulate[1|2]:LIMit:FAIL? <Boolean1>, … <Boolean5>

Returns:

<Boolean1> = summary, <Boolean2> = low-limit active, <Boolean3> = high limit active,

<Boolean4> = low-limit latched, <Boolean5> = high-limit latched


CALCulate:LIMit:LOWer[:POWer]

Description:

Set or return the lower limit power level for the selected channel. This limit is used for level
alarms. When the measured average power is below the lower limit, a down arrow ▼ will appear
on the display to the left of the measured value, and flag bits are set in the alarm register which
may be accessed using CALCulate:LIMit:FAIL? query and CALCulate:LIMit:CLEAR
commands.

Syntax:

CALCulate[1|2]:LIMit:LOWer[:POWer] <numeric value>

Argument:

<numeric value> = -99.99 to +99.99 dBm





Remote Operation

5-10

Advertising