9 output subsystem, Output:recorder:force, Output:recorder:max – Boonton 9240 RF Voltmeter User Manual

Page 71: Output:recorder:min, Output:recorder:source, 9 output subsystem -21, Output:recorder:force -21, Output:recorder:max -21, Output:recorder:min -21, Output:recorder:source -21

Advertising
background image

Boonton 9240 Series RF Voltmeter

5.5.9 OUTPut Subsystem

The OUTPut group of commands is used to control various outputs of the 9240. These outputs include the Recorder Output.
OUTPut commands for the Recorder Output include setting the DC output to the MAXimum or MINimum level or can be
forced to any level in between. Also the Recorder Output can be assigned to either channel 1 or 2.

OUTPut:RECorder:FORCe

Description:

Command sets the output voltage to the argument. Query returns the output voltage previously set
by FORCe or causes error -221 “Settings conflict” if the Recorder Ouput has been assigned to a
channel.

Syntax:

OUTPut:RECorder:FORCe <numeric value>

Argument:

<numeric value> = 0.000 to + 10.000 V

OUTPut:RECorder:MAX

Description:

Set or return the recorder output maximum, or full scale (+10.0V) voltage reference level.

Syntax:

OUTPut:RECorder:MAX

Argument: none


OUTPut:RECorder:MIN

Description:

Set or return the recorder output minimum, or downscale (0.0V) voltage reference level.

Syntax:

OUTPut:RECorder:MIN

Argument: none


OUTPut:RECorder:SOURce

Description:

Set or return the source channel for the Recorder Output.

Syntax:

OUTPut:RECorder:SOURce <character data>

Argument:

<character data> = CH1, CH2, ACTIVE