16 9230 emulation gpib commands, 16 9230 emulation gpib commands -40, Boonton 9240 series rf voltmeter – Boonton 9240 RF Voltmeter User Manual

Page 90: Remote operation 5-40

Advertising
background image

Boonton 9240 Series RF Voltmeter

5.5.16 9230 Emulation GPIB Commands

Table 5-2. 4230 EMULATION GPIB COMMANDS

Code

Description

Comment

AM

Measure A-B

AR

Measure

A/B

BD

Measure

B

-

A

BR

Measure

B/A

BN

4230A

Native

mode

CH

Channel select 1 - 2

CL

Clear

DB

dBm select

DF

Display

off

DN

Display

on

DR

dBr select

DU

Display

user

message

FA

Auto

filter

FL

Filter time select

0 to 20.00 in 0.05 steps

?ID

Talk

instrument

ID

*IDN?

Talk

instrument

ID

1

LH

High

limit

-99.99

to

99.99

in

0.01

steps

LL

Low

limit

-99.99

to

99.99

in

0.01

steps

LM0

Disable

limits

checking

function

LM1

Enable

limits

checking

function

LR Load

reference

PK0

Disable peak readout.

PK1

Peak sinewave voltage (x.1414 or +3dB)

VOLT

Volts units select (

µ

V, mV)

MF

Measure

filtered

MFD

Measure

Fast

Dual

Channels

MFS

Measure

fast

single

channel

MN

Measure

normal,

free

run

MS

Measure

settled

PW

Watts

select

RA

Autorange

RB

Recorder

bottom

Remote Operation

5-40

Advertising