Boonton 9240 Series RF Voltmeter


R T N

M O D E L

> 5 1 0 7 5

S E R

I A L

# 2 9 3 9 1

D A T E

0 8 2 3 1 0

U P S C A L E

0 : 5 0 0 0

D O W N S C A L E 0 : 0 0 0 0

M I

D S C A L E

0 : 0 0 0 0

Figure 3-14. Sensor Edit Data Menu Display



The parameter list will show TBLn (where n = 1, 2, 3, 4) when a serial number has not been entered for the corresponding
internal table. For example, TBL3 will be displayed if the serial number has not been previously entered for internal table 3. In
addition, the parameter list will show ADPTn (where n = 1, 2) if a serial number has not been entered for the table contained
within the sensor data adapter. For example, ADPT2 is displayed when the serial number has not been previously entered for
sensor data adapter 2.

To edit the voltage probe calibration data, move the cursor to the EDIT DATA function and press <Enter>. Scroll through the
voltage probe serial numbers until the desired selection is displayed. Press <Enter> to proceed. Enter the access code to edit or
depress the <Menu> key to escape. (See Figure 3-14.)

Sensor Edit Data Menu. An example of the Edit Data menu is shown in Figure 3-14. Table 3-7 contains a description of the
commands and associated parameters.

Table 3-7 Sensor Edit Data Menu Functions


Function Description

Parameters

Defaults


RTN

Returns the instrument to the

n/a

n/a


MODEL

Voltage probe model number

0 to 99999

0


SERIAL #

Voltage probe serial number

0 to 99999

0

DATE

Calibration date

MM/DD/YY

01/01/01

Where:

MM = 01 to 12

DD = 01 to 31

YY = 00 to 99


UPSCALE

Upscale linearity factors

Range : Factor

5000

[0 to 6] : [0 to 9999]


DOWNSCALE

Downscale linearity factors

Range : Factor

0

[0 to 6] : [-999 to 999]


MIDSCALE

Midscale linearity factors

Range : Factor

0

[0 to 6] : [-999 to 999]



Getting Started

3-16