background image

Boonton 9240 Series RF Voltmeter

STATus:QUEStionable:NTRansition

Description:

Set or return the value of the negative transition filter bit mask for the Questionable status group.
Setting a bit in the negative transition filter causes a 1 to 0 (negative) transition in the
corresponding bit of the Questionable Condition register to cause a 1 to be written in the
associated bit of the Questionable Event register.


Syntax: STATus:QUEStionable:NTRansition <numeric value>

Argument:

<numeric value> = 0 to 65535

Returns:

16-bit register value (0 to 65535)

STATus:QUEStionable:PTRansition

Description:

Set or return the value of the positive transition filter bit mask for the Questionable status group.
Setting a bit in the positive transition filter causes a 0 to 1 (positive) transition in the
corresponding bit of the Questionable Condition register to cause a 1 to be written in the
associated bit of the Questionable Event register.


Syntax:

STATus:QUEStionable:PTRansition <numeric value>

Argument:

<numeric value> = 0 to 65535

Returns:

16-bit register value (0 to 65535)

STATus:QUEStionable:CALibration:CONDition?

Description:

Return the current value of the Questionable Calibration Condition register. This register is used
to notify the user of questionable quality with respect to calibration. The following table shows
the bit assignments in the register. These bits are updated by the instrument in real time, and can
change in response to changes in the instrument’s operating condition.

Bit

Value

Definition

0

1

Not

Used

1

2

Not

used

2

4

Sens1 Default Shape

1 = Channel 1 using default shape table.

3

8

Sens2 Default Shape

1 = Channel 2 using default shape table.

Syntax:

STATus:QUEStionable:CALibration:CONDition?

Returns:

16-bit register value (0 to 65535)

Remote Operation