Chapter 10 troubleshooting, Chapter 10, Table 9 – Vaisala WMT52 User Manual

Page 111: Data validation, Troubleshooting

Advertising
background image

Chapter 10 __________________________________________________________ Troubleshooting

VAISALA______________________________________________________________________ 109

CHAPTER 10

TROUBLESHOOTING

This chapter describes common problems, their probable causes and
remedies, and includes contact information for technical support.

Table 9

Data Validation

Problem

Interpretation

Action

Wind measurement failure. Both
the speed and direction units are
replaced by a # sign or the data
values are irrelevant.

Blockage (trash, leaves,
branches, bird, snow, ice)
between the wind transducers.
Incorrect <cr><lf> settings in the
terminal program.

Remove the blockage, and
check that the wind transducers
are not damaged.
If the blockage is ice or snow, it
will melt after some time if
heating is enabled. Time for
clearance depends on the
severity of the weather event. If
birds are causing the blockage,
consider using the bird spike kit.

In ASCII and NMEA protocols
both <cr> and <lf> are required
after each command. Check that
your terminal program sends
both when pressing enter.
Note: The direction unit is # for
the wind speeds less than 0.05
m/s.