The following example shows how heating behaves as Ta starts to fall:

-

When Ta falls below +10 °C, heating is enabled.

-

Heating keeps Th > +4 °C until Ta < -11 °C.

-

Between -11 ... -65 °C, Th is approximately 15 °C warmer
than Ta.

When the heating function is disabled the heating is off in all conditions,
see

Supervisor Message on page 101

.

NOTE

Snow accumulation may cause a temporary wind measurement
problem even when the heating is enabled.