WIRING AND POWER MANAGEMENT

This chapter provides you with instructions on how to connect the
power supply and the serial interfaces, and how to manage and estimate
the average power consumption.

The WMT52 can be accessed through four different serial interfaces:
RS-232, RS-485, RS-422 and SDI-12. Each of them can be wired either
through the internal screw terminal or the 8-pin M12 connector
(optional). Only one serial interface can be used at a time.

Power Supplies

Operating Voltage

Operating voltage Vin+: 5 ... 32 VDC

Notice that for the average current consumption, see the graphs in

Figure 19 on page 40

. The minimum consumption graph is for SDI-12

standby mode.

The input power supply needs to be capable to deliver 60 mA (at 12 V)
or 100 mA (at 6 V) instant current spikes with duration of 30 ms. These

CAUTION

The cable openings in the sensor bottom assembly are covered with
hexagonal rubber plugs. If you are not using the cable glands (included
in the Bushing and Grounding Kit), keep the openings covered.