Appendix E ______________________________________________________ factory configurations

VAISALA______________________________________________________________________ 139

Supervisor Settings

0SF!0SF,t=19.8,b=17159,l=-50.0,m=4.0,h=10.0<cr><lf>

Table 21

General Unit Settings

Field Character

Field Name

Description

t

Heat control calibration temp.

-50.0...+60.0 °C (calibrate to Ta)

b

Direct ADC value of temp.diode

0...4096

l

Defrosting limit w. 50% duty
cycle

-100.0 ...[m] °C (default -4.0 °C)

m

Heating limit w. full power

[l]...[h] °C(default 0.0 °C)

h

Heating limit w. 50% duty cycle

[m]...100.0 °C (default 4.0 °C)