Table 11, Error messaging/text messages table – Vaisala WMT52 User Manual

Page 114

Advertising
background image

User’s Guide ______________________________________________________________________

112 __________________________________________________________________M210925EN-B

See also

Chapter 10, Troubleshooting, on page 109

. In case of constant

error, please contact Vaisala Service Center, see

Vaisala Service

Centers on page 108

.

Table 11

Error Messaging/Text Messages Table

Text message identifier (in
NMEA 0183 v3.0 protocol only)

Text Message

Interpretation and action

01

Unable to measure error

The requested parameters are
not activated in the message,
see

Chapter 8, Sensor and Data

Message Settings, on page 95

,

and check the parameter
selection fields.

02

Sync/address error

The device address in the
beginning of the command is
invalid. Ask the device address
with the ?! (SDI-12) or ?<cr><lf>
(ASCII and NMEA) command
and retype the command with
the correct address.

03

Unknown cmd error

The command is not supported,
use the correct command format,
see

Chapter 7, Getting the Data

Messages, on page 61

.

04

Profile reset

Checksum error in configuration
settings during power-up.
Factory settings used instead.

05

Factory reset

Checksum error in calibration
settings during power-up.
Factory settings used instead.

06

Version reset

New software version in use.

07

Start-up

Software reset. Program starts
from the beginning.

08

Use chksum xxx

Given checksum not correct for
the command. Use the proposed
checksum.

09

Measurement reset

The ongoing measurement of all
the sensors interrupted and
started from the beginning.

NOTE

When WMT52 is used on an RS-485 bus with other polled devices, the
error messaging feature shall always be disabled. This is done by the
following command: 0SU,S=N<crlf>.