b.2 Electrical Connections

b.2.1 Connecting Flow Transducers

Attach the 15 pin “D” connector cable(s) between the interface panel on the rear of the
SDPROC and the Flow Meter/Controller transducer(s). (Number of channels will vary
from one to four in accordance with configuration and options purchased.)

Figure b-1, Interface Panel for SDPROC.

No connection with the DATA IN/OUT, RELAYS or RS 232 connector is required for
operating the system. However, it may be desirable to perform remote monitoring or
control of the flow transducers. In this case, see the pin out chart for the “DATA
IN/OUT” connector in step b.2.4.

ƽ Important Notes:

Assure that the “Flow Transducer” cables are used in accordance with configuration
and options purchased.

In general, “D” Connector numbering patterns are standardized. There are, however,
some connectors with nonconforming patterns and the numbering sequence on your
mating connector may or may not coincide with the numbering sequence shown in our
pin configuration table above. It is imperative that the appropriate wires are matched in
accordance with the correct sequence regardless of the particular numbers displayed
on your mating connector, see the pin out chart for the “Flow Transducers” connector.

Power must be shut OFF when connecting or disconnecting any cables in the system.

Cable lengths for Flow Transducers and DATA IN/OUT may not exceed 9.5 feet (3 meters).