8 tor – maximum reading time, 9 vs- show parameter, 10 vsave – save parameter – AEG COMPACT READER ARE I2 User Manual

Page 20

Advertising
background image

Manual
Page 20 / 33- Revision 01

ARE I2 / RS 232 – Installation Guide

AEG
Identifikationssysteme GmbH

Söflinger Straße 100
D-89077 Ulm
Tel. ++49(0)731-933-1340/1877
Fax ++49(0)731-933-1855
e-mail: [email protected]

6.8 TOR – Maximum Reading Time

Timeout for the reader. TOR is used in operation mode 2 as maximum gating time for a

reading process . The length of the maximum gating time results from the equation gat-

ing_time = TOR * TB.

The time constant TB (Time Base) has always the default value 100ms.
Input format:

TOR <SP> parameter <CR>

Parameter:

Function

0

limits the reading process duration to exactly one reading cycle

1

limits the reading process duration to maximum 1 times TB

2

limits the reading process duration to maximum 2 times TB

...
255

limits the reading process duration to maximum 256 times TB

Output (example):

2 <CR>

6.9 VS- Show Parameter

The command VS lists all current parameter settings.

Input format:

VS <CR>

Output (example):

EC <SP> 0 <CR>
BD <SP> 2 <CR>

MD <SP> 2 <CR>

6.10 VSAVE – Save Parameter

All operating parameters temporarily stored are saved permanently using VSAVE.
Input format:

VSAVE <CR>

Output:

ok <CR>

Advertising