14.2. ETHERNET setting

ETHERNET can be configured in <P2.4 ETHERNET>.


Inventory of default parameters:

No NAME

VALUE

DESCRIPTION

P2.4.1

WORK MODE

SERVER, CLIENT

Ethernet connection as Server or Client.
SERVER – scale waits for connection
CLIENT – scale initiates the connection
to a HOST.

P2.4.2

IP ADDRESS

192.168.0.2

Setting an IP address.

P2.4.3

SUBNET MASK

255.255.255.0

Setting a subnet mask for Ethernet
connection.

P2.4.4

GATEWAY 192.168.0.1

Setting a gateway for Ethernet
connection.

P2.4.5

LOCAL PORT

4001

Setting a local port for Ethernet
connection. Only for devices that
work as SERVER. Servers waits
for connection on the specified port.

P2.4.6

HOST IP

192.168.0.3

Setting a host IP address (IP of a device
to connect with). Applicable only for
devices configured as CLIENTs.

P2.4.7

HOST PORT

2000

Setting a Host port (a port for connection
with a computer). Applicable only for
devices configured as CLIENTs.

P2.4.8

TIMEOUT 60

Time (in seconds) after which none-
active Ethernet connection is being
broken. Set to 0 to stop breaking the
connection.


Caution:

1. For appropriate setting of: <P2.4.2 IP ADDRESS>,

<P2.4.3 SUBNET MSK>, <P2.4.4 GATEWAY> contact
the supervisor of the net to connect with;

2. The way of connection via ETHERNET to the program

EDYTOR WPW is described in chapter 32.2.3.

3. The scale does not allow the automatic fetch of net

configuration from DHCP servers.