background image

Annually

Weekly

Daily

Calendar Time Using 8 Hour Running Time Rate Per Day

MTS

Cushion plates and wearable plates of the grip (molybdenum
disulfide grease)

MTS

Driven parts of lower crosshead (lubricating grease)

Frame, Integrated Operations Platform, and Work Area

MTS

Clean/replace air filter, suction filter, and high-pressure line filter

MTS

Clean off frame, Integrated Operations Platform, and work area

MTS

Change hydraulic oil

* There are no customer serviceable components on the MTS Criterion Series 60 systems. Maintenance
consists of keeping the frame and work area clean, general inspection, checking interlocks, and scheduled
frame calibration.

Warning

:

Detergents and cleaning fluids may be conductive.

Conductive fluids can cause circuits to short resulting in equipment damage.

Disconnect the power cord from the wall outlet before cleaning or inspecting any part of the
test frame. Ensure that the main power is off and the Electrical Disconnect is locked out. Be
careful not to spill any cleaning fluid on the frame.

Warning

:

Using the cleaning fluids incorrectly may cause injury to personnel or damage to the machine.

Check the cleaning fluids before using.

Follow all manufacture's recommendations and cautions when using any cleaning fluids.

General Cleaning

Clean your machine as often as needed. Use a damp, lint-free rag to clean the side covers, base, and
crosshead. If necessary, mild detergent or cleaning fluid can be used.

Monthly Maintenance

Verify that the Emergency Stop button is functioning properly. This can be done by pressing the Emergency
Stop button and making sure the machine stops and faults are generated, and then releasing the Emergency
Stop button.

Verify any additional interlocks are functioning properly (for example, the travel limit switch on the frame).