For example, the rebar specimen used in this example test could tolerate installation forces up to 60
kN before permanent deformation would result. Because the Running Maximum/Minimum meter reads
only .554 kN tension and 1.123 kN compression at the conclusion of specimen installation, it is safe to
say that the specimen has not been damaged.

2. Change the load meter to timed type.

Right-click on the Load meter and select Meter Type > Timed. This meter will show you in real time
the load on the specimen as you run the test.

3. Zero the actuator (displacement) meter.