Chapter 4 Setup Menus

23

4.2 Communications Port Setup Menus


These menus allow you to access the communications port options that the
PM296/RPM096

uses to communicate with a master computer or a printer. Table

4-2 lists the communications options, their code names and applicable choices.

From the main menu, select the menu for the port you want to configure. Activate
the middle window to scroll through the list of available options, and then activate
the lower window to set the option value.

To select and view a setup option:

Press to activate the middle window.

Use the up/down arrow keys to scroll to the desired option. The option setting will

appear in the lower window.

To change the selected option:

Press to activate the lower window.
Press the up/down arrow keys to scroll to the desired value.
Press to store the selected value or press to quit the setup menu.

NOTE:

An optional analog expander can be connected to Communications Port #2.

Table 4-2 Communications Options

(

∗ default setting

)

COM1

Code Parameter Options Description

Prt.1

Communications
protocol

ASCII

rtu
dnP3
Prnt

ASCII protocol
Modbus RTU protocol
DNP3.0 protocol
Printer mode

rS

Interface standard

232

RS-232 interface

485

RS-485 interface

Addr

Address

0

-99 ASCII

1

-247 Modbus

0*-255 DNP3.0

Powermeter address

bAud

Baud rate

110

110 baud

300

300 baud

600

600 baud

1200

1200 baud

2400

2400 baud

4800

4800 baud

9600

9600 baud

19.20

19,200 baud

38.40

38,400 baud

Prt.1

ASCII

Prot

ESC

ENTER

SELECT

SELECT

SELECT

CHG

ENTER

Prt.1

ENTER

SELECT

CHG

ENTER

Prt.2

ENTER