Rockwell Automation 1441-DYN25-MRUC Run Up Coast Down Extension Module for the Dynamix 2500 User Manual

Page 32

Advertising
background image

32

Rockwell Automation Publication 1441-UM006A-EN-P - May 2011

Chapter 2

Run Up/Coast Down Extension Module

Spacing

Controls the spacing between collected spectra:

Δ

rpm is a fixed speed change between spectra.

%

Δ

rpm is a percentage speed change between spectra.

Δ

Time is a fixed time interval between spectra.

Δ

rpm +

Δ

Time is a combination of speed and time interval settings. If

speed stops changing then spectra are calculated using the Time Interval
value.

%

Δ

rpm +

Δ

Time is a combination of speed and time interval settings.

If speed stops changing then spectra are calculated using the Time
Interval value.

Δ

rpm

%

Δ

rpm

Δ

Time

Δ

rpm +

Δ

Time

%

Δ

rpm +

Δ

Time

Your selection affects the Speed Interval
and Time Interval settings.

Speed Interval (RPM) Sets the speed interval used in calculating FFT spectra.

This parameter not used when Spacing is set to

Δ

Time.

1…240 rpm

Time Interval (sec)

Sets the time interval used in calculating FFT spectra.

This parameter is not used when Spacing is set to

Δ

RPM and %

Δ

RPM.

.001…604800 s

Display Units

Sets the Y-axis units.

The available units depends on the transducer used to collect data.

See

Run Up/Coast Down Setup Parameters on page 18

.

Detection

Sets the signal detection and scaling display for the plot:

Peak - Measured RMS value multiplied by the square root of two (1.4142).
Pk-Pk - Measured RMS value multiplied by two times the square root of

two (2.8284).

RMS - Root mean squared signal level of the input signal.

RMS

Peak

PkPk

Lines

The number of lines (bins) of resolution for the FFT spectra. The larger the
number of lines, the better the frequency resolution of the measured data in the
spectra. Similarly, the higher the number of lines, the longer it takes to display
the spectra in the plot.

100

200

400

800

1600

3200

6400

12800

X-axis

Sets the x-axis frequency units for displaying spectra/FFT data:

Hz - Displays spectra in Hertz, cycles per second.
CPM - Displays spectra in cycles per minute.
Orders - Displays spectra in orders of the operating speed.

Hz

CPM

Orders

Your selection affects the Freq Range and
LF Cutoff parameters.

Freq. Range (EU)

The maximum frequency for the measurement.

The frequency range is determined by the X-axis parameter.

10…2000 Hz

600…120000 CPM

.001…166 Orders

Table 7 - Waterfall Plot Parameters

Parameter Name

Description

Values/Comments