Spectrogram plot parameter descriptions on – Rockwell Automation 1441-DYN25-MRUC Run Up Coast Down Extension Module for the Dynamix 2500 User Manual

Page 37

Advertising
background image

Rockwell Automation Publication 1441-UM006A-EN-P - May 2011

37

Run Up/Coast Down Extension Module Chapter 2

Table 9 - Spectrogram Plot Parameter Descriptions

Parameter Name

Description

Values/Comments

Start Spd (RPM)

Sets the start speed of the recorded data for the plot.

This defaults to the Start Spd value set in the RUCD Setup screen.

See

Run Up/Coast Down Setup Parameters on page 18

.

10…3600 rpm

Stop Spd (RPM)

Sets the stop speed of the recorded data for the plot.

This defaults to the Stop Spd value set in the RUCD Setup screen.

See

Run Up/Coast Down Setup Parameters on page 18

.

10…3600 rpm

Spacing

Controls the spacing between collected spectra:

Δ

rpm is a fixed speed change between spectra.

%

Δ

rpm is a percentage speed change between spectra.

Δ

Time is a fixed time interval between spectra.

Δ

rpm +

Δ

Time is a combination of speed and time interval settings. If

speed stops changing then spectra are calculated using the Time
Interval value.

%

Δ

rpm +

Δ

Time is a combination of speed and time interval

settings. If speed stops changing then spectra are calculated using the
Time Interval value.

Δ

rpm

%

Δ

rpm

Δ

Time

Δ

rpm +

Δ

Time

%

Δ

rpm +

Δ

Time

Your selection affects the Speed Interval
and Time Interval settings.

Speed Interval (RPM)

Sets the speed interval used in calculating FFT spectra.

This parameter not used when Spacing is set to

Δ

Time.

1…240 rpm

Time Interval (sec)

Sets the time interval used in calculating FFT spectra.

This parameter is not used when Spacing is set to

Δ

RPM and %

Δ

RPM.

.001…604800 s

Display Units

Sets the Y-axis units.

The available units depends on the transducer used to collect data.

See

Run Up/Coast Down Setup Parameters on page 18

.

Detection

Sets the signal detection and scaling display for the plot:

Peak - Measured RMS value multiplied by the square root of two

(1.4142).

Pk-Pk - Measured RMS value multiplied by two times the square root of

two (2.8284).

RMS - Root mean squared signal level of the input signal.

RMS

Peak

PkPk

Lines

The number of lines (bins) of resolution for the FFT spectra. The larger the
number of lines, the better the frequency resolution of the measured data in
the spectra.

Similarly, the higher the number of lines, the longer it takes to display the
spectra in the plot.

100

200

400

800

1600

3200

6400

12800

X-axis

Sets the X-axis frequency units for displaying spectra/FFT data:

Hz - Displays spectra in Hertz, cycles per second.
CPM - Displays spectra in cycles per minute.
Orders - Displays spectra in orders of the operating speed.

Hz

CPM

Orders

Your selection affects the Freq Range and
LF Cutoff parameters.